HB Isolator

Technical data: HB-ISOLATOR-2W-1C

2-wire sensors like the pH sensor (HBPH-Mk2) are sensitive to grounding and in areas with electromagnetic interference you might need an isolator. The symptom is a surprisingly steady measurement and typically more than one sensor providing the same output.

This isolator PR3186A is loop powered and easy to install between the PLC/measuring device and the sensor.  The isolator makes the sensor circuit independent of other ground signals.

This isolator is only suited for the two-wire sensors. You can read more about the isolator here.

Product features

  Tested and reliable
5 years warranty
Easy to install

Applications

HB Isolator Application

Specifications

Power supply
Power supply6 - 35 V
Power dissipation, per channel50 mW (3186A) Vterminal x I (3186B)
Isolation voltage, test / working2.5 kVAC / 300 VAC (reinforced)
Zone 2 / Div. 2250 VAC
Response time (0...90%, 100...10%)< 5 ms
AccuracyBetter than 0.05%
Measurement range3.5...23 mA
2-wire transmitter supply3.5...32.5 V
Signal conversion1:1
Signal range3.5...23 mA
Signal range, input to output3.8...20.5 mA
Output loop current limitation, typ.24 mA
Current output overload, max.50 mA
Output3.5...23 mA
Connection2-wire
Environmental conditions
Installation inPollution degree 2 & meas. / overvoltage cat. II
Operating temperature-25...+70°C
Storage temperature-40...+85°C
Calibration temperature20...28°C
Protection degreeIP20
Mechanical specifications
Dimensions (HxWxD)113 x 6.1 x 115 mm
Weight approx.70 g
DIN rail typeDIN EN 60715/35 mm
Wire size0.13...2.5 mm2 / AWG 26...12 stranded wire
Screw terminal torque0.5 Nm
VibrationIEC 60068-2-6
2...25 Hz±1.6 mm
25...100 Hz±4 g
Approvals
ATEXKEMA 10ATEX0147 X
IECExKEM 10.0068X
UKEXDEKRA 21UKEX0055X
c FM usFM17US0004X / FM17CA0003X
c UL us, UL 61010-1E314307
CCC2020322310003554
EAC ExRU C-DK.HA65.B.00355/19
DNV MarineTAA00001RW

Ordering code

Ordering code
HB-ISOLATOR-2W-1C